| Display |
51 mm x 32 mm, dual function LCD display, 15 mm (0.6") digit size |
| Measurement |
pH (0 to 14 pH)
MV (-1999 to 1999 mV) |
| Temperature Compensation for pH Measurement |
Manual –0 to 100 C, able to be adjusted by pushing a button on the front panel |
| pH Calibration |
pH7, pH4 and pH10, 3 points of calibration to ensure the best linearity and accuracy |
| Data Hold |
Hold the current reading on the display |
| Memory Recall |
Maximum and Minimum readings can be saved and retrieved by the Recall function |
| Data Output |
RS-232 PC Serial Interface |
| pH Electrode |
Any pH electrode with a BNC connector |
| Sampling Time |
Approximately 0.8 seconds |
| Power Supply |
9 VDC Battery, Approximately 7mA |
| Weight |
250g / 0.55 lb. (battery included) |
| Included Accessories |
pH electrode (760pH)
ATC Temperature Probe (760ATC)
Conductivity Probe (760CP)
Dissolved Oxygen Probe (760DOP)
Software and Communication Cable (AK 760) |
| Electrical Specifications (232 + -5 °C) |
| pH |
Range: 0 to 14 pH
Resolution: 0.01 pH
Accuracy: + -(0.02 pH + 2 dgt) |
| mV |
Range: 0 to 1999 mV
Resolution: 1 mV
Accuracy: + -(0.5% + 2dgt) |
* pH accuracy is based on calibrated meter only.
* Specification tests under the environment RF Field Strength less than V/M & frequency less than 30 MHz only |