Home > Calibrators >

Fluke Calibration 9144A Field Metrology Dry-Well Calibrator



 
Our Price: $5,150.00

Quantity in Stock:1
Product Code: 9144A
Qty:

Description
 

Features

  • Field Metrology Well, 50 °C to 660 °C, w/9144-INSX
  • Temperature Range at 23 °C : 50 °C to 660 °C (122 °F to 1220 °F)
  • Display Accuracy: ± 0.35 °C at 50 °C ± 0.35 °C at 420 °C ± 0.5 °C at 660 °C
  • Stability: ± 0.03 °C at 50 °C ± 0.05 °C at 420 °C ± 0.05 °C at 660 °C
  • Axial Uniformity at 40 mm (1.6 in): ± 0.05 °C at 50 °C ± 0.35 °C at 420 °C ± 0.5 °C at 660 °C
  • Radial Uniformity: ± 0.02 °C at 50 °C ± 0.05 °C at 420 °C ± 0.10 °C at 660 °C
  • Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes): ± 0.015 °C at 50 °C ± 0.025 °C at 420 °C
  • ± 0.035 °C at 660 °C
  • Hysteresis: 0.1°C
  • Operating Conditions: 0 °C to 50 °C, 0 % to 90 % RH (non-condensing)
  • Environmental Conditions (for all specifications except temperature range): 13 °C to 33 °C
  • Immersion (Well) Depth: 150 mm (5.9 in)
  • Insert OD: 24.4 mm (0.96in)
  • Heating Time: 15 min: 50 °C to 660 °C
  • Cooling Time: 35 min: 660 °C to 50 °C 25 min: 660 °C to 100 °C
  • Resolution: 0.01 °
  • Display: LCD, °C or °F user-selectable
  • Size (H x W x D): 290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in)
  • Weight: 17 lbs (7.7kg)
Each Unit Includes...
  • Insert A, B, C, D, E or F
  • RS-232 and 9930 Interface-it control software

These units are packed with functionality and are remarkably easy to use. Field Metrology Wells are lightweight, small, and quick to reach temperature set points; yet they are stabile, uniform, and precise. This industrial product line is perfect for transmitter-loop calibration, comparison calibration, or a simple check of a thermocouple sensor. With the “process” option, you have no need to carry additional tools into the field. It includes a built-in readout for resistance, voltage, and mA measurement, 24V loop power and on-board documentation.


Fluke Calibration 9142/9143/9144 Specifications

9142/9142-P

9143/9143-P

9144/9144-P

Temperature Range (at 23° C Ambient)

–25° C to 150° C

(-15° F to 302° F)

33° C to 350° C

(91° F to 662° F)

50° C to 660° C

(122° F to 1220° F)

Display Accuracy

±0.2° C (full range)

±0.2° C (full range)

±0.35° C at 50° C

±0.35° C at 420° C

±0.5° C at 660° C

Ref. Readout Accuracy (Process version)

±0.02 to ±0.04° C

±0.025 to ±0.055° C

±0.025 to ±0.09° C

Stability

±0.01° C (full range)

±0.02° C at 33° C

±0.02° C at 200° C

±0.03° C at 350° C

±0.03° C at 50° C

±0.05° C at 420° C

±0.05° C at 660° C

Radial Uniformity

±0.01° C (full range)

±0.01° C at 33° C

±0.015° C at 200° C

±0.02° C at 350° C

+/-0.02 °C at 50 °C

+/-0.05 °C at 420 °C

+/-0.1 °C at 660°C

Axial Uniformity
(bottom 1.6 inches)

±0.05° C (full range)

±0.04° C at 33° C

±0.1° C at 200° C

±0.2° C at 350° C

±0.05° C at 50° C

±0.35° C at 420° C

±0.5° C at 660° C

Immersion Depth

6 in

Heating Time (max)

23 min

5 min

15 min

Cooling Time

15 min to –25° C

14 min to 100° C

25 min to 100° C

Operating Conditions

0° C to 50° C, 0 % to 90%RH (non-condensing)

Size (HWD)

11.4 x 7.3 x 11.6 in

Weight

18 lbs

RTD Inputs

2-, 3-, 4-wire RTD. Resistance range of 0 to 400 ohms. Ni-120, PT-100 (385), PT-100 (3926), PT-100 (JIS), or ohms

TC Inputs

Type J, K, T, E, R, S, L, N, and mV

Transmitter Range

0-24 mA with 24-28 VDC loop power

mA Accuracy

0.02% of rdg + 0.002 mA